- Page 1
- Page 2 - Page 3 - Page 4 - Page 5 - Page 6 - Page 7 - Page 8 - Page 9 - Page 10 - Page 11 - Page 12 - Page 13 - Page 14 - Page 15 - Page 16 - Page 17 - Page 18 - Page 19 - Page 20 - Page 21 - Page 22 - Page 23 - Page 24 - Page 25 - Page 26 - Page 27 - Page 28 - Page 29 - Page 30 - Page 31 - Page 32 - Page 33 - Page 34 - Page 35 - Page 36 - Page 37 - Page 38 - Page 39 - Page 40 - Page 41 - Page 42 - Page 43 - Page 44 - Page 45 - Page 46 - Page 47 - Page 48 - Page 49 - Page 50 - Page 51 - Page 52 - Page 53 - Page 54 - Page 55 - Page 56 - Page 57 - Page 58 - Page 59 - Page 60 - Page 61 - Page 62 - Page 63 - Page 64 - Page 65 - Page 66 - Page 67 - Page 68 - Page 69 - Page 70 - Page 71 - Page 72 - Page 73 - Page 74 - Page 75 - Page 76 - Page 77 - Page 78 - Page 79 - Page 80 - Page 81 - Page 82 - Page 83 - Page 84 - Page 85 - Page 86 - Page 87 - Page 88 - Page 89 - Page 90 - Page 91 - Page 92 - Page 93 - Page 94 - Page 95 - Page 96 - Page 97 - Page 98 - Page 99 - Page 100 - Page 101 - Page 102 - Page 103 - Page 104 - Page 105 - Page 106 - Page 107 - Page 108 - Page 109 - Page 110 - Page 111 - Page 112 - Flash version © UniFlip.com |
T H YROID UPTA K E SYST EM
SPECIFICATIONS:
MEDICAL SPECTROMETER HARDWARE All-In-One Flat Panel PC: 15.6" Color Touchscreen, Windows 10 Enterprise LTSC, Solid State Hard Drive, Ethernet, USB, Video/Audio Out, Built-In Speakers and Color Printer. Multichannel Analyzer: Channels: 1024 Spectral Resolution: FWHM 8% Count Rate: (Maximum) 150,000 cps for Tc-99m Count Rate Stability: 99% Gross Count Rate Linearity: Within 5% up to 150,000 cps Pulse Height Linearity: Within 2% (independent of detector) Detector High Voltage Adjustment: Automatic adjustment for both probe and well; uses 10 µCi Cs-137 as the calibration source. MEDICAL SPECTROMETER SOFTWARE Radionuclides: Factory Programmed: Au-198, Ba-133, Co-57, Co-58, Co-60, Cr-51, Cs-137, F-18, Fe-59, Ga-67, Ga-68, Gd-153, Ge-68/Ga-68, Hg-197, I-123, I-125, I-131, In-111, Ir-192, K-42, Lu-177, Na-22, Na-24, Pd-103, Ra-223, Se-75, Sr-85, Tc-99m, Tc-99m/Tl-201, Tl-201, Yb-169, Wide Window User Set: Unlimited user defined isotopes, setting ROI, half life, name, efficiency and range. OTHER HARDWARE: Probe: 2" x 2" Nal (Tl) integral line scintillation detector Uptake Stand: Dimensions: Arm down and retracted: 34" l x 26" w x 55" h (86.4 x 66 x 139.7 cm) Maximum Height: 60" (152.4 cm) Maximum Length: Arm Extended: 61" (154.9 cm) Collimated Shield: Flat field collimator meeting IAEA specifications Positioning: Distance bar and positioning light Arm: Counterbalanced, two section arm, moves 24.5" (62.2 cm) vertically and extends 31" (78.7 cm) horizontally from stand's vertical column. Casters: 4" Total locking Weight: 280 lb (127 kg) Combined weight with Well Counter: 330 lb (150 kg) Power: Low voltage transformer with hospital grade power cord and plug. Line Voltage: 120 VAC, 60 Hz, 2 AMP circuit breaker 230 VAC, 50 Hz, 1 AMP circuit breaker OPTIONAL: 187-602 Well Counter: Detector: 2" x 2" Nal (Tl) integral line scintillation detector with a .75" dia x 1.44" deep well (1.9 x 3.7 cm) Lead Shielding: 1" thick (2.5 cm) Cover: .125" thick (.32 cm) Weight: 50 lb (22.7 kg) 187-603 Lead Shield, Well Counter: Detector: 2" x 2" Nal (Tl) Lead Shielding: 1" thick (2.5 cm) Weight: 69 lb (31.4 kg) Certification: See website for details. Warranty: Two-years parts and labor
PRINTED REPORTS
Wipe Test Report
Printed on: 1/31/2013 10:28:11 AM
Full Spectrum
Net Counts Total Net CPM Test Time (sec) : 3,656 : 7,260.0 : 30
Wide Window
Total Efficiency Background Net Activity Net CPM ROI (KeV) Trigger Level : 35.90 % : 949 DPM : 19,768 DPM : 6,872.0 : 25 - 750 : 20,000 DPM
Net Counts > 750 keV : 52 Test Date Area Name Location Status : 1/17/2013 5:22:35 PM : Hot Lab : counter
Performed By Detector Net CPM 0 DPM 143 DPM 60 DPM 42 DPM 18 DPM 8 DPM 6 DPM 6,872.0 2,198.0 1,130.0 1,872.0 1,362.0 422.0 382.0
: thomas simpson : Well - 123 (S/N: 123)
: FAIL - An Isotope Trigger Level
Isotope Background 949 DPM 226 DPM 75 DPM 286 DPM 131 DPM 37 DPM 34 DPM MDA
Net Activity 19,768 DPM 5,769 DPM 1,247 DPM 14,871 DPM 4,689 DPM 650 DPM 470 DPM
Trigger
Status
Wide Window Ba-133 (Well) Co-57 Cs-137 F-18 I-123 Tc-99m
20,000 DPM Pass 2,000 DPM Fail 20,000 DPM Pass 20,000 DPM Pass 20,000 DPM Pass 2,000 DPM Pass 20,000 DPM Pass
Comments :
Signature :
Software Ver: 1.0.57.0 Page 1 of 1
Date Signed :
Wipe Test Report
Daily Calibration Details Report Full-size sample reports can be viewed at www.biodex.com/thyroiduptake/reports
187-600 Thyroid
Uptake System, Atomlab 960, 120 VAC, Mobile, PC • All-in-one flat panel PC featuring Windows® Operating System, Atomlab 960 PC software, printer and 1024 channel multichannel analyzer • 2" x 2" tube assembly and base • Mobile support stand with collimator Uptake System, Atomlab 960, 230 VAC, Mobile, PC • All-in-one flat panel PC featuring Windows® Operating System, Atomlab 960 PC software, printer and 1024 channel multichannel analyzer • 2" x 2" tube assembly and base • Mobile support stand with collimator Interface, Atomlab 960 DICOM Well Counter, 1" Lead (2.5 cm) Includes cover and calibration fixture. Lead Shield, Well Counter, 1" Lead (2.5 cm) Rod Source, Cs-137, Calibrated, 0.1 µCi Check Source, Cs-137, 10 µCi * Uncalibrated, 1" dia x .25" thick (2.5 x .64 cm) Thyroid Uptake Neck Phantom (Complete with bottle carrier, capsule holder and 12 polyethylene bottles)
187-601 Thyroid
Optional:
187-146 187-602 187-603
Related:
063-139 101-103 043-365
Thyroid Uptake Neck Phantom – designed to simulate a patient’s neck. See page 89 for details.
*Recommended Check Source for calibration of probe.
Place an Order or Receive a Quote sales@biodex.com • +1-631-924-9000 • 1-800-224-6339